What is the conductivity limit for cleaning validation?

Published in Cleaning Validation Limits 2 mins read

The conductivity limit sometimes proposed for rinse water samples in cleaning validation is ≤2.4 µS/cm, especially when the rinse water temperature is around 65°C.

While a conductivity limit is sometimes proposed for rinse water in cleaning validation, understanding what this value measures is crucial. Here's a breakdown of using conductivity in cleaning validation:

Understanding Conductivity in Cleaning Validation

Conductivity measures a solution's ability to conduct an electrical current. In the context of cleaning validation, it indicates the presence of ionic species (e.g., salts, acids, bases) in the rinse water. A higher conductivity value generally suggests a greater concentration of these substances.

Typical Conductivity Limit

  • A typical value sometimes proposed for rinse water samples in cleaning validation protocols is ≤2.4 µS/cm.
  • This value is sometimes proposed when the rinse water is near a temperature of 65°C.

Factors Affecting Conductivity Measurement

Several factors can influence the conductivity reading:

  • Temperature: Conductivity is temperature-dependent; higher temperatures generally result in higher conductivity readings. This is why specifying the temperature (e.g., 65°C) is important when establishing a conductivity limit.
  • Water Quality: The initial conductivity of the water used for rinsing (e.g., purified water, water for injection) must be considered. Background conductivity will contribute to the overall reading.
  • Residue Type: Different residues will have varying conductivities. For instance, ionic compounds will significantly increase conductivity, while non-ionic compounds may have a minimal impact.

Considerations When Using Conductivity as a Cleaning Validation Criterion

  • Specificity: Conductivity is a non-specific measurement. It doesn't identify the specific residue present but indicates the overall ionic burden.
  • Correlation: Establish a correlation between conductivity and the concentration of the target residue. This can be achieved by spiking studies where known amounts of the residue are added to the rinse water, and the corresponding conductivity is measured.
  • Alternative Methods: Consider using more specific analytical methods (e.g., HPLC, TOC) in conjunction with conductivity to provide a more comprehensive assessment of cleaning effectiveness.