Why is Field Diaphragm Used?

Published in Microscopy Illumination 3 mins read

The field diaphragm is a crucial component in microscopy, primarily used to control the illumination pathway and improve image quality.

At its core, the field diaphragm controls how much light enters the substage condenser and, consequently, the rest of the microscope. This function is essential for optimizing the light path and preventing unwanted light from reaching the objective lens and eyepiece.

The Role of the Field Diaphragm

Think of the field diaphragm as an adjustable aperture located near the light source. By opening or closing this diaphragm, you control the area of the illuminated field in the specimen plane. This is different from the aperture diaphragm (usually located within or below the condenser), which controls the angle of the light rays entering the objective.

Key Reasons for Using the Field Diaphragm

Controlling the amount of light entering the condenser has several important benefits:

  • Reduces Glare: Excess light outside the field of view can scatter within the microscope optics, causing glare and reducing contrast. By closing the field diaphragm so that the illuminated area just covers the field of view visible through the eyepiece, this scattered light is minimized.
  • Improves Contrast: By eliminating peripheral light scatter, the field diaphragm helps to increase the contrast of the specimen, making fine details more visible.
  • Optimizes Illumination: It allows you to match the illuminated area to the field of view for a specific objective lens. When you switch to a higher magnification objective (which sees a smaller area), you should also close the field diaphragm to match the smaller field of view.
  • Aids in Condenser Alignment (Kohler Illumination): Proper use of the field diaphragm is a fundamental step in setting up Kohler illumination, a technique used to produce even, bright, and glare-free illumination of the specimen. In Kohler illumination, the field diaphragm is closed down to a small opening and then focused by adjusting the height of the condenser. The condenser is then centered using the visible image of the diaphragm's edge.

Practical Application

When using a microscope, after placing your slide and focusing on the specimen, you should typically:

  1. Open the field diaphragm fully.
  2. Close the field diaphragm slowly until its edge is just visible in the field of view.
  3. Adjust the height of the condenser to bring the edge of the field diaphragm into sharp focus.
  4. Center the condenser using the centering screws until the diaphragm image is in the center of the field of view.
  5. Open the field diaphragm just enough so that its edge is no longer visible, but not much wider.
  6. Adjust the aperture diaphragm (usually below the condenser) to fine-tune contrast and resolution.

By following these steps and correctly using the field diaphragm, you ensure that only the area of the specimen you are currently observing is illuminated, leading to significantly better image quality.