The concentration activity is calculated by adjusting for non-ideal behavior using activity coefficients. However, the direct question relates to activity concentration in the context of nuclide detection. Therefore, activity concentration is calculated using detector-specific calibration.
Calculating Activity Concentration
Activity concentration, specifically in the context of detecting radionuclides, is calculated using the following method:
-
Adjusted Net Peak Areas: Begin with the measured peak areas in a spectrum, adjusted to account for background noise and any necessary corrections.
-
Calibration Factor: Multiply the adjusted net peak areas by a detector- and nuclide-specific calibration factor. This factor accounts for the detector's efficiency in detecting the specific nuclide and its decay characteristics. According to the reference, "The activity concentration is calculated from the adjusted net peak areas by multiplication with a detector- and nuclide-specific calibration factor".
Formula
The general formula is:
Activity Concentration = Adjusted Net Peak Area * Calibration Factor
Example
Imagine you're measuring the activity concentration of Cesium-137 (137Cs) using a gamma-ray spectrometer.
- Adjusted Net Peak Area: Let's say the adjusted net peak area for the 137Cs peak at 662 keV is 1000 counts.
- Calibration Factor: Assume the detector's calibration factor for 137Cs is 0.01 Bq/count (Becquerels per count).
Then, the activity concentration would be:
Activity Concentration = 1000 counts * 0.01 Bq/count = 10 Bq
This means the activity concentration of 137Cs in the sample is 10 Becquerels.