How Do You Measure the Amount of Protein in a Sample?

Published in Protein Analysis 4 mins read

One of the simplest ways to measure the amount of protein in a sample is by directly measuring absorbance at 280 nm using a spectrophotometer. However, several other methods are also commonly employed, each with its advantages and disadvantages depending on the specific application, required accuracy, and the characteristics of the protein sample.

Common Methods for Protein Quantification

Here's a breakdown of common protein quantification methods:

  • UV Absorbance at 280 nm:

    • Principle: Proteins containing aromatic amino acids (tryptophan, tyrosine, and phenylalanine) absorb UV light maximally at 280 nm. The absorbance is directly proportional to the protein concentration.
    • Advantages: Simple, rapid, non-destructive (the sample can be used for further experiments).
    • Disadvantages: Accuracy can be affected by the presence of nucleic acids (which also absorb at 280 nm) and other UV-absorbing substances. Also, the method requires knowledge of the protein's extinction coefficient for accurate quantification. Proteins lacking aromatic amino acids will not be accurately measured.
  • Bradford Assay:

    • Principle: This colorimetric assay is based on the binding of the dye Coomassie Brilliant Blue G-250 to proteins, causing a shift in the dye's absorption spectrum. The amount of bound dye is proportional to the protein concentration.
    • Advantages: Relatively simple, rapid, and sensitive.
    • Disadvantages: The assay's response can vary depending on the specific protein being measured, requiring the use of a protein standard (like bovine serum albumin or BSA) for calibration. Certain detergents can interfere with the assay. The dye binds primarily to basic and aromatic amino acids.
  • Lowry Assay:

    • Principle: This colorimetric assay involves a reaction between protein and Folin-Ciocalteu reagent, resulting in the development of a blue color.
    • Advantages: More sensitive than UV absorbance at 280 nm.
    • Disadvantages: Time-consuming, sensitive to interfering substances (e.g., reducing agents, detergents), and exhibits protein-to-protein variability. Requires careful control of reaction conditions.
  • Bicinchoninic Acid (BCA) Assay:

    • Principle: Proteins reduce Cu2+ to Cu+ in an alkaline medium. BCA then reacts with Cu+ to form a purple-colored product that absorbs strongly at 562 nm.
    • Advantages: More tolerant to detergents than the Bradford assay, good sensitivity, and less protein-to-protein variability compared to the Lowry assay.
    • Disadvantages: Can be affected by reducing agents and lipids.
  • Amido Black Assay:

    • Principle: This assay is based on the binding of Amido Black dye to proteins. The unbound dye is then removed and the bound dye is solubilized, followed by measuring the absorbance of the solution.
    • Advantages: Works well with membrane proteins as the dye is less impacted by detergents.
    • Disadvantages: The assay takes about an hour to complete.
  • Kjeldahl Method:

    • Principle: This is a classical method that measures the total nitrogen content of a sample. The protein concentration is then estimated based on the assumption that protein contains a relatively constant percentage of nitrogen (approximately 16%).
    • Advantages: Accurate for determining total protein in complex samples.
    • Disadvantages: Time-consuming, requires specialized equipment, and does not provide information about specific proteins. It also measures non-protein nitrogen, which can lead to overestimation.
  • Amino Acid Analysis:

    • Principle: This method involves hydrolyzing the protein into its constituent amino acids and then quantifying each amino acid using chromatography.
    • Advantages: Highly accurate, provides detailed information about the amino acid composition of the protein.
    • Disadvantages: Expensive, time-consuming, and requires specialized equipment and expertise.

Choosing the Right Method

The best method for measuring protein concentration depends on the specific application and sample characteristics. Factors to consider include:

  • Accuracy requirements: For high accuracy, amino acid analysis or Kjeldahl method may be preferred.
  • Sample purity: UV absorbance is suitable for relatively pure protein samples. Colorimetric assays are more versatile for complex samples.
  • Presence of interfering substances: Some assays are more tolerant to detergents or reducing agents than others.
  • Sample volume: Some assays require larger sample volumes than others.
  • Equipment availability: UV spectrophotometers are readily available, while other methods may require specialized equipment.